Table 2 Error in actual osteotomy vs. planned osteotomy including Cadaver 1 (with Grey resin) and Cadaver 2 Right (with Biomed Amber resin) and Cadaver 2 Left (with Tough 2000 resin)

From: The feasibility of a novel 3D-Printed patient specific cutting guide for extended trochanteric osteotomies

 

Longitudinal Osteotomy

Transverse Osteotomy

Planned Length (mm)

Actual Length

(mm)

Percent Error

(mm)

Planned Width

(mm)

Actual Width

(mm)

Percent Error

(mm)

Cadaver 1

144.25

143.0

0.86%

27.26

28.0

2.71%

Cadaver 2 Left

134.67

131.5

2.35%

12.46

13.3

6.86%

Cadaver 2 Right

141.22

136.6

3.27%

7.75

7.5

3.23%