Table 1 Test–retest reliability of DICOM measurements. DICOM landmark measurements of four cases that were performed a second time by the same observer with an interval of one month between the measurements. mean m1: mean of first measurement, mean m2: mean of second measurement, range min: lowest difference between the two measurements, range max: highest difference between the two measurements, range mean: average difference between the two measurements ICC: intra class correlation coefficient, CI: confidence interval

From: Accuracy of pelvic bone segmentation for 3d printing: a study of segmentation accuracy based on anatomic landmarks to evaluate the influence of the observer

 

mean

range

 

ICCa 95% CI

landmark

m1

m2

min

max

mean

ICCa

lower limit

upper limit

LM1 [mm]

290.2

291.0

0.6

3.4

1.7

0.995

0.947

1.000

LM2 [°]

124.7

124.2

0.8

3.7

2.6

0.984

0.789

0.999

LM3 [mm]

232.3

231.1

2.2

3.7

2.9

0.980

0.808

0.999

LM4 [mm]

112.2

114.2

0.1

3.6

2.0

0.984

0.592

0.999

LM5 [mm]

120.2

121.4

0.6

2.4

1.6

0.988

0.806

0.999

LM6 [mm]

94.9

93.6

0.1

8.3

2.8

0.887

0.092

0.992

LM7 [mm]

107.4

110.3

2.3

9.5

4.7

0.881

0.240

0.992

  1. aTwo-way mixed effects, absolute agreement, single rater/measurement