Skip to main content

Table 1 Physical X-ray shielding properties of different filaments. Values are the measured remaining radiation values detected below the printed discs in Gy/min

From: Custom-made 3D-printed X-ray shield for tumor-specific irradiation of xenograft mice

Setting

ADC,

200 kV, 20 mA,

Shelf position 2,

Co 0.5 mm filter

ADC,

160 kV, 20 mA,

Shelf position 2,

Co 0.5 mm filter

160mV, 20 mA,

Shelf position 2

no filter

Shield material

No shield

0.951

0.55

10.225

Polylactic acid (PLA)

0.926

-

-

Copper

0.407

0.193

0.064

Bronze

0.379

0.178

0.051

Tungsten

0.018

0.007

0.008

Aluminum

0.892

0.515

1.555

Steel

0.402

0.193

0.064

  1. # The values without filter may deviate from the measurements with filter, as the device has an internal sensitivity setting for the selected filter. The values can only be compared within each column